Display Hold - Uline Ohaus Adventurer H-5276 Manual Del Usuario

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

DISPLAY HOLD

Two modes are available: (See Figure 58)
Display Hold - allows the user to capture and store a
stable weight.
Peak Hold - allows the user to capture and store the
highest stable weight.
Display Hold
Place sample on the pan. Press Hold to begin.
0.00
>0<
Gross:
0.00 g
Mode:
Tare:
0.00 g
Automatic Mode:
Item
Hold
Settings
Figure 58
DISPLAY HOLD
1.
In the upper left portion of the home screen, select
Display Hold. (See Figure 59)
2. Place the sample on the pan and press HOLD at
any time while the weight is being captured.
3. The main display line now shows the first stable weight.
4. Press CLEAR to remove the hold and return to Display
Hold home screen.
Figure 59
Display Hold
Gross:
Tare:
Item
Settings
PAGE 15 OF 54
OPERATION CONTINUED
The
Max 4200 g
d=0.01 g
DISPLAY HOLD
home screen
Main Display Line
g
Display Hold
Reference Fields
Off
Save to
Function Buttons
Menu
USB
Max 4200 g
Stable weight is curently being held.
436.64
436.64 g
Mode:
0.00 g
Sample Name:
Save to
Cancel
USB
PEAK HOLD
1.
In the upper left portion of the home screen, select
Display Hold. (See Figure 60)
2. Choose Peak Hold mode in item settings.
(See below)
3. Place sample on the pan and press START to begin.
4. Continue to weigh samples. The highest stable
weight will be held.
5. To remove the hold and return to normal operation
press STOP.
Figure 60
ITEM SETTINGS
To view or adjust the current settings, touch the
ITEM SETTINGS button. The settings screen appears.
(See Figure 61)
Figure 61
d=0.01 g
Mode: Choose between Peak Hold and Display Hold
g
(default).
Display Hold
Sample Name: Assign a name to the sample.
Menu
Print Settings: Change printing settings.
NOTE: The SAVE TO USB button will only appear
after a USB flash drive has been connected to
the balance.
Display Hold
Highest stable weight is curently being held.
475.64
Gross:
475.64 g
Mode:
Tare:
0.00 g
Sample Name:
Item
Save to
Stop
Settings
Display Hold
Mode
Sample Name
Print Settings
Back
Max 4200 g
d=0.01 g
g
Peak Hold
Menu
USB
Exit
0521 IH-5276

Publicidad

Tabla de contenido
loading

Tabla de contenido