CLEANING CONTINUED
REMOVING AND REINSTALLING THE GLASS DOORS FOR CLEANING:
1. On the back of the balance,
press the pin and slide out the
door. (See Figure 68)
Figure 68
OPERATING ISSUE
Balance will not turn on
Poor accuracy
Cannot calibrate
Cannot change menu settings
Low Reference weight
Invalid Piece Weight
Operation Timeout
Screen Displays "------"
If the troubleshooting section does not resolve your problem, contact ULINE Customer Service at 1-800-295-5510.
PAGE 18 OF 54
MAINTENANCE CONTINUED
2. After cleaning, slide the doors
into the slot while pressing the pin
mentioned in step one.
(See Figure 69)
TROUBLESHOOTING
CAUSES
No power to balance
Improper calibration
Unstable environment
Calibration menu locked
Approved Mode set to ON
Unstable environment
Incorrect calibration masses
Sub-menu locked
Approved Mode set to ON
Reference weight too small
The weight on the pan is too small to define a valid
reference weight
Average piece weight is too small
Weight reading is not stable
Busy (tare, zero, printing, waiting for a stable weight)
1-800-295-5510
Figure 69
Verify connection and voltage
Platform calibration
Move balance to suitable location
Turn Calibration menu lock off
Turn Approved Mode off
Move balance to suitable location
Use correct calibration masses
Unlock sub-menu
Turn Approved Mode off
Increase sample size
Increase average piece weight
Move balance to suitable location
Wait until completion
uline.com
3. Slide the doors into the slot until
the back stopper aligns with the
other door. (See Figure 70)
RECOMMENDATIONS
Figure 70
0521 IH-5276