Specifications

Specifications are subject to change without notice.
Specification
Measurement method
Dimensions (W x D x H)
Weight
Enclosure
Protection class
Installation category
Power requirements
Operating temperature
Storage temperature
Humidity
Display
Optical light source
Measurement units
Range
Accuracy
Linearity
1
http://www.hach.com
2
1 mNTU = 0.001 NTU
Details
Nephelometry with the scattered light collected at a 90° angle to the incident light
and 360° around the sample vial.
41 x 28 x 12.5 cm (16 x 11 x 7.7 in.)
2.37 kg (5.23 lb)
Instrument: III; Power supply: II
2
II
Instrument: 15 VDC, 2 A; Power supply: 100 to 240 VAC, 50/60 Hz
10 to 40 °C (50 to 104 °F)
-30 to 60 °C (-22 to 140 °F)
5 to 95% relative humidity, non-condensing
17.8 mm (7 in.) color touch screen
Class 2 laser product: Contains a non user-serviceable class 2 laser.
NTU, FNU, TE/F, FTU, EBC, mg/L, mNTU
0 to 700 NTU, FNU, TE/F, FTU; 0 to 100 mg/L; 0 to 175 EBC
± 2% of reading plus 0.01 NTU from 0 to 40 NTU
± 10% of reading from 40 to 700 NTU based on Formazin primary standard at
25 °C (77 °F)
Better than 1% for 0 to 40 NTU on Formazin at 25 °C (77 °F)
1