Agilent Technologies TwisTorr 305-IC Manual De Instrucciones página 284

Tabla de contenido

Publicidad

8
Technical Information
Window Description
WIN #
READ
WRITE
370
Always
-
371
Always
-
372
Always
-
373
Always
-
374
Always
-
375
Always
-
376
Always
-
377
Always
-
382
Always
-
384
Always
-
399
Always
-
400
Always
-
401
Always
-
402
Always
-
404
Always
-
406
Always
-
407
Always
-
503
Always
Always
724
Always
Always
725
Always
Always
Below is the table with the Admitted Values:
WIN #
MIN
MAX
0 VALUE
0
Stop
1
Off
8
Serial
100
Disable
101
0
3
102
0
99999
103
0
999999
104
High Level
105
0
100
108
0
6
284/302
TYPE
NAME
N
N
N
N
N
N
N
N
Average Pump
N
Temperature
N
Average Pump Power
N
Controller life
A
Program Listing CRC
A
Bootloader CRC
A
Parameter Listing CRC
Parameter Listing
A
Struct.CRC
Program Listing code
A
and revision
Parameter Listing code
A
and revision
N
RS485 Address
N
Run Up time
L
RUTimeCtrl
1 VALUE
UNIT
Start
-
On
-
Remote
-
Enable
-
-
Hz/W/s
s
Low Level
-
%
-
TwisTorr 305-IC User Manual / 87-901-058-01 (A.00)
MEANING
Worst temp cycle power
Worst temp cycle frequency
Worst power cycle number
Worst power cycle time
Worst power cycle temperature
Worst power cycle current
Worst power cycle power
Worst power cycle frequency
It indicates the average pump temperature
It indicates the average supply power
It indicates the total hours of controller life
It is the CRC of the actual embedded program listing
It is the CRC of the embedded bootloader program
It is the CRC of the actual embedded parameter listing
It defines the structure of Parameter Listing accepted by embedded
Program Listing
It defines the structure of Program Listing accepted by embedded
Program Listing
It indicates code and release of embedded Parameter Listing
It defines the serial address in RS485 communication (accessible with
every serial address)
Run Up time
Run Up time control
WIN #
MIN
MAX
109
110
111
0
2
117
120
121
122
125
0
2
126
0
65535
143
0
1
DEFAULT
(FACTORY
SET)
---
---
---
---
---
---
---
---
---
---
---
---
---
---
---
---
---
0
15
On
0 VALUE
1 VALUE
UNIT
-
Reset
-
Impulse
Continuous
-
-
Hz
Hz
Hz
Off
On
-
-
0,2s
-