Startup And Operation Of The Twistorr 305-Ic - Agilent Technologies TwisTorr 305-IC Manual De Instrucciones

Tabla de contenido

Publicidad

Startup and Operation of the TwisTorr 305-IC

To turn on the TwisTorr 305-IC via the on board controllers, supply the pump
with the proper supply voltage and then act on the "START"and "INTERLOCK"
signals. Once these signals are recognized the controller will start the pump.
Never connect the user cable to the DB15 connector on the IC system or
disconnect it from said DB15 connector with the user cable powered.
WARNING!
Before carrying out any operation to connect to or disconnect from the DB15
connector on the IC system, you must first make sure that the user cable is not
powered (power unit must be switched off).
When disconnecting after switching off the IC system, wait until all the LEDs are
switched off before disconnecting the user cable from the system.
If this warning is not respected the IC system, or devices connected to it, could be
damaged.
The user cable can be disconnected from the IC system while the user cable
powered exclusively in the case of an emergency stop.
(1) In order to find the location of the supply voltage input pins (on the IN/OUT
NOTES
and Power Supply Connector), as well as the input voltage admitted values,
please refer to "Technical Specifications" section and to "IN/OUT and Power
Supply" section of this instruction manual.
(2) In order to find the location of the START/STOP and INTERLOCK input pins
(on the IN/OUT and Power Supply Connector), as well as the related
electrical ratings, please refer to the "IN/OUT and Power Supply" section of
this instruction manual.
(3) In order to find details about how to connect the START/STOP and the
INTERLOCK signals, please refer to the "How to Connect the Open Collector
Inputs of the Controller" section of this instruction manual.
(4) The TwisTorr 305-IC is provided with a DB15 mating connector, not pre-wired.
This DB15 mating connector is suitable to withstand to the current level
needed to supply the controller and it is fully compatible with the DB15
connector which is mounted on the control unit. It can be used, after properly
wiring, both to connect the TwisTorr 305-IC to the supply source as well as to
the accessory parts (cooling fan, vent valves) and also to the serial comm.