Figure 3-1 Input Power Connection P1 (Power Input/Fuse Cover Shown Removed) - Siemens SITRANS F FUS1020 IP65 NEMA 4 Instrucciones De Servicio Resumidas

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Commissioning
2.2 Mounting the Flowmeter
I/O Terminals TB1 and TB2
Input Power Connector P1
Fuse F1 Type GMA, 2A
Power Input/Fuse Cover
Power Cables
Figure 3-1
Input Power Connection P1 (Power Input/Fuse cover shown removed)
7. Replace Power Input/Fuse Cover and tighten setscrew.
8. Connect the power cables to the appropriate power source (100-240 VAC @ 50/60 Hz or
Note
The branch circuit must be limited to 20A. It is recommended that the circuit breaker be
located near the flowmeter.
9. If unit is operational, turn power off, replace Keypad Enclosure cover using six #8
12
9-36 VDC) and power up unit.
securing screws (torque to 7 oz-in). Reinstall transducer cables.
Data/Control Cables
1012CN Series Transducer Cables
Fill boots with CC110 to ensure weather tight
connection (supplied)
J1 RS-232 Port DB9 Connector
Keypad Enable Switch
SITRANS FUS1020 IP65 NEMA 4 Quick Start
Operating Instructions (Compact), 4/2009, CQO:QSG006 Revision B

Publicidad

Tabla de contenido
loading

Tabla de contenido