Transducer Selection - Example for D1H High Precision Transducer
1. At the [Chan/Path Setup] menu, trigger the<Down Arrow> to select [Pick/install Xdcr].
2. Trigger <Right Arrow> then <Down Arrow> to [Transducer Model].
3. Trigger <Right Arrow> sensor.
4. Trigger <Down Arrow>and scroll to [1011 Hi Precision]. Trigger <ENTER> to select.
5. Trigger <Down Arrow> to select [Transducer Size].
6. Trigger <Right Arrow> sensor.
7. Trigger <Down Arrow> and scroll to [D1H]. Trigger <ENTER> to select.
8. IMPORTANT: Record Number Index and Spacing Method. Trigger <Down Arrow> as
9. Transducers can now be mounted.
Transducer Identification
The transducer part number located on the front face provides a detailed identification. For
example, the
FUS1010 IP65 NEMA 4X Quick Start
Operating Instructions, 12/2008, CQO:QSG002 Revision B
needed to locate data.
Select based on type.
Select based on size.
After transducer is mounted select "Install."
Part Number: 1011HNFS-D1HUT1-S2 means:
Model
Hi Precision
NEMA w/F-Conn
Pipe Material
3.4 Setting the Parameters
Agency Approved
Temperature
Size
Commissioning
27