Philips T Serie Instrucciones De Uso página 10

Ocultar thumbs Ver también para T Serie:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 39
Sensor Cleaning and Low-Level Disinfection
The reusable sensors should be cleaned and disinfected, but never
sterilized. Follow the procedure outlined below.
Warnings
Use only validated cleaning agents and disinfectants listed below; do
not use any others. Failure to do so may damage the sensor or its con-
necting wires and shorten product lifetime or cause safety hazards.
Select disinfectants carefully as some have very similar names but
completely different compositions.
Do not immerse the sensor connector in any of the cleaning solutions,
disinfectant, or other liquid (only the sensor and cable housing may be
immersed, not the connector).
Do not soak sensors in disinfectants longer than specified by the dis-
infectant manufacturer.
Do not sterilize the sensors.
Validated Cleaning Agents
- Mild Detergent
Validated Disinfectants
®
Metricide
28
®
Metricide
Plus 30
®
Terralin
Liquid
®
Incidin
Liquid
Step
1
Clean sensor following instructions supplied with cleaning agent.
2
Disinfect sensor following instructions supplied with disinfectant.
3
Rinse sensor in water, wiping it dry with a clean cloth, leaving to dry com-
pletely. If you notice any deterioration or damage to sensor or cable, dis-
pose the sensor immediately.
- Salt Solution (1%)
®
Cidex
Formula 7
®
Cidex
OPA
®
Cidex
Plus
®
Omnicide
Cleaning and Low-Level Disinfection
Kohrsolin
Mucapur
Isopropanol (70%) or
Isopropanol Wipe (70%)
28
8
®
(2%)
®
-CD (1%)

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

M1191tM1192tM1193t

Tabla de contenido