15 Examine the transducer and cable for damage, such as cracks or splitting where fluid can enter.
If damage is evident, discontinue use of the transducer, and contact FUJIFILM SonoSite or your local
representative.
B
Option
Clean and disinfect system and transducer to a low level
(non-critical uses)
Use the following procedure to clean and disinfect the ultrasound system and transducer if it has not come
into contact with blood, broken skin, mucosal membranes, or bodily fluids.
WARNING
Follow the manufacturer's instructions when using cleaners and disinfectants. The cleaners and disinfectants
listed in the procedure are both chemically compatible and have been tested for efficacy with the system
and transducers. Confirm that the cleaners and disinfectants are appropriate for your facility's use.
To avoid electrical shock, before cleaning, disconnect the system from the power
WARNINGS
Wear the appropriate personal protective equipment (PPE) recommended by the
Do not spray cleaners or disinfectants directly on the system surfaces or on
Caution
Use only FUJIFILM SonoSite-recommended cleaners and disinfectants. Using a
Do not attempt to disinfect a transducer or transducer cable using a method or

Clean and disinfect system and transducer to a low level (non-critical uses)

If the system or transducer has come into contact with any of the following, use
the high-level cleaning and disinfection procedure. See
system and transducer to a high level (semi-critical uses)"
Blood
Broken skin
Mucosal membranes
Bodily fluids
supply.
chemical manufacturer, such as eyewear and gloves.
system and transducer connectors. Doing so may cause solution to leak into the
system, damaging it and voiding the warranty.
non- recommended disinfecting solution or incorrect solution strength can
damage the System and transducer and void the warranty. Follow the disinfectant
manufacturer's recommendations for solutions strengths.
chemical that is not included here. Doing so can damager the transducer and void
the warranty.
"Clean and disinfect
on page 7: