Electrode Use and Care
WARNING: This product should be used only under the direct supervision of a
registered healthcare practitioner. Product should be used according to the in-
structions in this manual. Improper use of this or any electrode may result in a
patient burn. Due to the potential for causing skin burns, devices must be limited
to 0.25W per cm2. Electrodes should not be placed on the body so that the Avid
IF's electrical current flows across the chest. This may cause cardiac arrhythmia.
NOTE: The electrodes supplied with this Avid IF2 have been designed
specifically for use with this product. They must not be manipulated, cut,
or altered in any way. This may affect the even distribution of stimulation
and/or its intensity beyond the recommended 0.25W per cm
NOTE: Do not place electrodes on cut, broken or irritated skin. If irritation
occurs, consult with your physician or therapist.
To apply self-adhesive, reusable electrodes:
^ Attach the electrodes to the lead wires provided with the Avid IF2.
^ Grasp an electrode by the edge and gently peel away the electrode
from its plastic liner.
^ Apply the electrodes to the treatment area by firmly pressing the
electrode against the skin. Check for proper adhesion. Your physician
or therapist will show you where to apply the electrodes.
^ Always make sure electrodes are placed at least one inch apart
from each other.
To remove the electrodes:
^ Grasp the electrode by the edge and gently remove it from the skin.
Return it to the plastic liner.
^ Replace the electrodes on the plastic liner in the resealable pouch
provided.
CAUTION: Never remove the electrode by pulling on the wire. This
may damage your electrode.
CAUTION: Be careful when using the Avid IF2 while sleeping.
Wires may tangle and become loose.
16 | Avid IF2 Patient Manual
2
maximum.