Technical Data & Specifications (Cont.)
Pulse coordination modes:
Continuous: Stimulation is provided at the user-selected beat frequency.
1|1:
After one second of stimulation, the beat frequency abruptly changes from
60% of user setting to 160% of user setting. One second later, the beat
frequency abruptly changes back to 60% of user setting. The cycle then
starts over.
6|6:
After six seconds of stimulation, the beat frequency abruptly changes from
60% of user setting to 160% of user setting. Six seconds later, the beat
frequency abruptly changes back to 60% of user setting. The cycle then
starts over.
6/6:
The beat frequency gradually changes from 60% of user setting to 160% of
user setting over a six-second period. Over the next six seconds, the beat
frequency gradually changes back to 60% of user setting. The cycle then
starts over.
Equipment Classification and Device
Symbol Descriptions
Symbol
20 | Avid IF2 Patient Manual
Meaning
Stimulator, lead wires and electrodes
are type BF Applied parts
Indicates the manufacturer
Indicates the manufacturer's serial
number
Refer to instructional manual
Warning symbol