Instructions
To Assemble the Handpiece (continued)
1. Using aseptic technique, remove the irrigation tube spike cap from the
handpiece irrigation tubing and install the irrigation tube spike into the irrigation
bag aseptic peel port (figure 1).
NOTES:
• Using a regulated facility vacuum source and a fluid collection canister with
the SurgiLav Plus Irrigation System is optional and depends on the SurgiLav
Plus handpiece configuration and tubing assembly.
• If using suction, make sure a fluid collection canister is connected properly
to a regulated facility vacuum source per local hospital protocol. Adjust the
regulated facility vacuum source per local recommendations/regulations
governing procedure-specific suction levels.
• The suction tube adapter may be cut off to connect the handpiece suction
tubing to a smaller diameter port on the fluid collection canister if necessary.
• To facilitate assembly, pull away and separate the irrigation tubing from the
co-bonded suction tubing enough to allow easy access to the corresponding
connection points.
2. Connect the handpiece suction tubing to the fluid collection canister, if
required.
NOTE: The suction tube adapter may be removed to connect the handpiece
suction tubing to a smaller diameter suction port on a vacuum fluid collection
canister.
3. Install the appropriate tip assembly into the handpiece until an audible 'click'
is heard. See the Accessories section for tip assembly information.
Figure 1 – To Assemble the Handpiece
4
Print Date: Aug 29, 2016 01:59:30 PM
0000004063, Rev. L Effective Date: -
To Operate Handpiece
WARNINGS:
• ALWAYS follow the current local recommendations and/or regulations
governing infection control protocols for pulsatile lavage treatments. Failure to
comply may result in the transmission of dangerous bacteria into the air and
onto surfaces from wounds.
• ALWAYS place tubing sets away from personnel traffic areas to eliminate
inadvertent tubing set kinking and/or handpiece movement.
• ALWAYS use aseptic technique when placing the handpiece in a safe and
secure location when not in use.
• ALWAYS operate the equipment within the specified environmental condition
values. See the Specifications section.
• DO NOT place the handpiece on a patient.
• If using suction, ALWAYS follow current local recommendations/regulations
governing procedure-specific suction levels to remove fluid waste safely from
a surgical site.
1. Squeeze the trigger enough to allow fluid to reach the end of the handpiece
tip assembly before operating.
2. To start irrigation, squeeze the trigger.
NOTE: If desired, push the trigger lock to hold the trigger in the high or low speed
irrigation setting. To disengage the trigger lock, move the trigger lock to the unlock
position.
3. If using suction, slide the suction clamp to open and close the suction tubing
pathway to control suction. If the handpiece fails to provide suction, see the
Troubleshooting section.
4. To stop irrigation and end the operation of the handpiece, release the trigger.
To Disassemble the Handpiece
WARNINGS:
• DO NOT reuse single use, disposable accessories. All disposable accessories
are single use only.
• Handling biohazard waste is potentially dangerous. ALWAYS follow current local
regulations governing biohazard waste to safely handle and dispose of surgical
fluid waste.
• A tip assembly and suction tubing will contain surgical waste after use.
ALWAYS handle these disposable accessories as potentially infectious
materials after use. ALWAYS wear gloves and protective eye wear when
removing and disposing of these disposable accessories.
NOTE: Make sure all the surgical fluid waste has been removed from the tip
assembly and suction tubing before disconnecting the suction tubing from the
fluid collection canister.
1. Press the latch mechanism on the tip assembly and remove the tip assembly
from the handpiece (figure 2).
Figure 2 – To Remove the Tip Assembly
0207-050-702 Rev-L
www.stryker.com