EN
Drill and Reamer Attachments
1. Slide the retaining collar back.
2. Hold the retaining collar in position and insert the tool
into the attachment.
3. Release the retaining collar.
4. Gently pull on the tool to verify it is secure.
5. To remove the tool, slide the retaining collar back
again.
12
To Operate a Handpiece
WARNINGS:
▪ DO NOT open the handpiece housing during surgery
as the inside of the device is not sterile.
▪ ALWAYS follow the specified duty cycle to prevent the
equipment from overheating. See the Specifications
section.
▪ ALWAYS consult attachment and tool labeling for
product-specific duty cycles and instructions for use.
▪ DO NOT use this equipment in areas in which
flammable anesthetics or flammable agents are mixed
with air, oxygen, or nitrous oxide.
▪ Take special precautions regarding electromagnetic
compatibility (EMC) when using this equipment. Place
the equipment into service according to the EMC
information contained in this manual.
▪ Portable and mobile radio frequency (RF)
communications equipment can affect the function of
this equipment.
▪ This equipment should not be used adjacent to or
stacked with other equipment. If adjacent or stacked
use is necessary, ALWAYS verify normal operation in
the configuration in which it will be used.
▪ ALWAYS operate the equipment within the specified
environmental condition values:
Temperature
Humidity
Limitation
Limitation
1. Before operating the equipment, verify the following:
▪ You are using Stryker-approved equipment.
▪ The equipment is not damaged.
▪ The equipment is clean and sterile.
▪ You have properly assembled the equipment.
▪ You have installed a fully-charged power module.
NOTE: If you experience difficulty during operation,
contact your Stryker sales representative or call Stryker
customer service. See the Contact Information section.
Outside the US, contact your nearest Stryker subsidiary.
7305-001-700 Rev-A
Atmospheric
Pressure Limitation
www.stryker.com