PRODUCT DESCRIPTION

Based on Oscillometric methodology and silicon integrated pressure sensor, blood pressure and pulse rate
can be measured automatically and non-invasively. The LCD display will show blood pressure and pulse rate.
The most recent 60 measurements can be stored in the memory with date and time stamp. The Electronic
Sphygmomanometers corresponds to the below standards: IEC 60601-1:2005/EN 60601-1:2006/AC:2010
(Medical electrical equipment – Part 1: General requirements for basic safety and essential performance),
IEC60601-1-2:2007/EN 60601-1-2:2007 /AC:2010 (Medical electrical equipment – Part 1-2: General
requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility
- Requirements and tests), IEC 80601-2-30:2009+Cor.2010/EN 80601-2-30:2010 (Medical electrical equipment
– Part 2-30: Particular requirements for the basic safety and essential performance of automated non-invasive
sphygmomanometers),EN 1060-1: 1995 + A1: 2002 + A2: 2009 (Non-invasive sphygmomanometers - Part 1:
General requirements), EN 1060-3: 1997 + A1: 2005 + A2: 2009 (Non-invasive sphygmomanometers - Part
3: Supplementary requirements for electro-mechanical blood pressure measuring systems). ANSI/AAMI
SP-10:2002+A1:2003+A2:2006
SPECIFICATIONS
Product name: Blood Pressure Monitor
Model: BPW-1005
Classification: Internally powered, Type BF applied part, IPX0, No AP or APG, Continuous operation
Machine size: Approx. 85mm x 64.5mm x 28mm (3 11/32"x 2 17/32"x 1 3/32")
Accuracy: Pressure: 5ºC-40ºc within±0.4kpa (3mmHg) pulse value:±5%
Cuff circumference: 14cm - 19.5cm (5 1/2" - 7 11/16")
Weight: Approx. 110g (3 7/8 oz.) (exclude batteries)
Measuring method: Oscillometric method, automatic inflation and measurement
Memory volume: 60 times with time and date stamp
Power source: Batteries: 2 × 1.5V
Measurement range: Cuff pressure: 0-300mmHg
Systolic: 60-260mmHg | Diastolic: 40-199mmHg
Pulse rate: 40-180 beats/minute
Accuracy: Pressure: ±3mmHg | Pulse rate: ±5%
Environmental temperature for operation: 5ºC~40ºC (41ºF~104ºF)
Environmental humidity for operation: ≤90%RH
Environmental temperature for storage and transport: -20ºC~55ºC (-4ºC~131ºC)
Environmental humidity for storage and transport: ≤90%RH
Environmental pressure: 0kPa-105kPa
Battery life: Approx 270 times
All components belonging to the pressure measuring system, including accessories: Pump, Valve, LCD, Cuff, Sensor
Note: These specifications are subject to change without notice.
Size AAA
4 4
NOTIC
1. Read
2. Stay
3. The c
4. Durin
5. Meas
6. Plea
arm
for lo
7. Cons
1) Th
2) Th
sh
3) Th
4) Si
5) Ne
8.
youn
9. Do n
10. Bloo
obse
Natio
11. Infor
mon
ELEC
12. If Irre
press
Sphy
cons
Ther
1) Th
2) Th
pe
13. Plea
haza

Publicidad

Tabla de contenido
loading

Tabla de contenido