4.9
Danger of burns
Argon plasma can strongly heat electrically conductive materials,
e.g. metal. To avoid burns, never direct argon plasma towards
electrically conductive materials.
4.10 Risk of accidental tissue damage with the APC 300
If the gas bottles of the APC 300 are not opened or are com-
pletely empty, a spray coagulation is performed with the instru-
ment after each activation. There is an audible warning signal;
the bottles on the display of the APC 300 blink and an error mes-
sage appears. In the spray coagulation mode there is a danger
of unintentional tissue damage; the instrument can also be dam-
aged!
Before using the APC 300 always open the gas bottles. Observe
the self-check performed by the unit. How much gas is there left
in the bottles? The bottle symbols on the display of the APC 300
give an approximate indication of the quantity of gas remaining.
Never work with two gas bottles which are almost completely
empty. When one gas bottle is empty, the APC 300 switches over
to the second gas bottle automatically.
Always observe the warning signals and error messages.
4.11 Modification of the product by the user
Erbe Elektromedizin expressly warns against modifying the
product. Any modification exempts Erbe Elektromedizin from
any and all liability.