Erbe 20132-046 Manual Del Usuario página 25

Cable de conexión apc
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 62
OBJ_DOKU-189884-007.fm Seite 25 Freitag, 11. Juni 2021 10:59 10
The detergent and disinfectant must be suitable for medical de-
vices made of plastic and metal and have a pH value between 5.5
and 11.
Observe the specifications from the manufacturer of the deter-
gent and disinfectant and in the chapter "Overview of validated
procedures".
1. Prepare a cleaning bath and a separate disinfectant bath.
2. Immerse the product completely in a cleaning bath, ensuring
that the product does not come into contact with other com-
ponents in the bath.
3. Thoroughly clean the surfaces with a soft brush/soft dispos-
able cloth.
4. Completely fill a disposable syringe with cleaning solution
and rinse the lumina (at least 5 times).
5. Rinse the surfaces with sterile, deionized water (at least
1 minute).
6. Completely fill a disposable syringe with sterile, deionized
water and rinse the lumina (at least 5 times).
7. Repeat the preceding cleaning steps if the product or used
rinse water still contains visible contamination.
8. Immerse the product completely in a disinfectant bath, en-
suring that the product does not come into contact with oth-
er components in the bath.
9. Completely fill a disposable syringe with disinfecting solution
and rinse the lumina (at least 5 times).
10. Rinse the surfaces with sterile, deionized water (at least
1 minute).
11. Completely fill a disposable syringe with sterile, deionized
water and rinse the lumina (at least 5 times).
12. Remove the irrigation adapter.
13. Dry the product until there are no more fluid residues visible
(e.g. with filtered compressed air).