OBJ_DOKU-163748-015.fm Seite 25 Donnerstag, 10. Dezember 2020 1:48 13
1. Connect the product to the multifunctional socket of
VIO 300 D or VIO 3.
Note: You can also connect the product to a MF-2 socket of a
VIO 300 D unit or a MF-U socket of a VIO 3 using the 20183-101
adapter.
6.5
Conducting a performance test
1. Moisten a gauze strip with a sterile NaCl solution.
2. Grasp the gauze strip with the instrument jaws and activate
the HF current.
Steam will form after approx. 2 to 3 seconds if the instrument is
working properly.
If damaged, do not use this product!
6.6
Using the product
The instrument is ready for use with 5 mm trocars. To insert into
a trocar sleeve, the instrument jaws must be closed.
To close the jaws, press the handle together. To open the jaws,
pull the handle apart.
ERGO handle (20195-240): The upper part of the handle can be
rotated freely relative to the lower part. The required rotational
position can be set using the locking knob.
Do not coagulate unless the contact surfaces are within your
field of vision. Do not touch any metallic instruments during this
process.
6.7
Precleaning the product in the procedure room
1. Use a soft cloth (such as a compress) to remove any obvious
contamination immediately after use and/or place the prod-
uct in deionized water.
25