3.5.10 Install modules
Add modules for additional output communications options. Refer to the documentation that is
supplied with the module.
3.6 Plumbing
3.6.1 Connect the drain lines
Chemical exposure hazard. Dispose of chemicals and wastes in accordance with local,
regional and national regulations.
Connect the supplied
For analyzers with an enclosure, refer to
For analyzers without an enclosure, refer to
Note: Analyzers without enclosures do not have a case drain.

3.6.2 Drain line guidelines

Incorrect installation of the drain lines can cause liquid to go back into the instrument and cause
damage.
• Make sure that the drain lines are open to air and are at zero back pressure. Refer to
• Make the drain lines as short as possible.
• Make sure that the drain lines have a constant slope down.
• Make sure that the drain lines do not have sharp bends and are not pinched.
Figure 13 Drain lines open to air
1 Sample drain tubing

3.6.3 Sample line guidelines

Select a good, representative sampling point for the best instrument performance. The sample must
be representative of the entire system.