2.5.1 Status indicator light

The status indicator light shows the status of the analyzer. Refer to
is above the display.
Color
Status
Green The analyzer is in operation with no warnings, errors or reminders.
Yellow The analyzer is in operation with active warnings or reminders.
Red
The analyzer is not in operation due to an error condition. A serious problem has occurred.
2.6 Items to collect
Collect the items that follow to install the instrument. The items that follow are supplied by the user.
In addition, collect all of the personal protective equipment appropriate to the chemicals that are
handled. Refer to the current safety data sheets (MSDS/SDS) for safety protocols.
Section 3 Installation
Multiple hazards. Only qualified personnel must conduct the tasks described in this
section of the document.
3.1 Installation guidelines
Install the analyzer:
• Indoors in a clean, dry, well-ventilated and temperature-controlled location.
• In a location with minimum mechanical vibrations and electronic noise.
4
Use fasteners applicable to the mounting surface (¼-in. or 6 mm SAE J429-Grade 1 bolts or
stronger).
Table 5 Status indicator description
Fasteners to
mount the
analyzer on a
wall if
applicable
4
(4x)
Drill
Deionized
water (or
sample
water)
3M KCl
electrolyte,
150 mL
Level
Wire strippers
0.5M Sodium
Nitrate, 500 mL
Diisopropylamine,
99%, 1 L (or
Ammonia 28%, 1 L)
C A U T I O N
Table
5. The status indicator light
Tape
measure
Wire cutters
10-mg/L
Sodium
Chloride
Standard,
500 mL
100 µm
filter for
each
sample line
(optional)