OBJ_DOKU-193153-006.fm Seite 28 Mittwoch, 25. März 2020 4:12 16
7
Information on electromagnetic
compatibility (EMC) of the units
Notwithstanding the EMC specifications in the ERBEJET 2 and
VIO user manuals, HF emissions are classified as Class A
cording to CISPR 11 for common use of the units for the system
apply. All other EMC specifications in the user manuals of the in-
dividual units also persist for common use of the units.
8
How to use
8.1
Checking the expiration date
The use-by-date is indicated on the packaging. Do not use this
product if the use-by-date has been exceeded!
8.2
Checking and opening the packaging
Only use this product if the packaging is unopened and undam-
aged. If the packaging is opened or damaged, the product is no
longer sterile and must not be used.
Use aseptic techniques when removing the product from the
packaging!
8.3
Checking the product
Check the product for damage before use, including all the insu-
lation.
If damaged, do not use this product!
1.) The system is suited for use in environments other than domestic
areas and in ones directly connected to a puplic power supply sys-
tem that also supplies buildings being used for domestic purposes.
The customer or the user of the system should ensure that it is used
in such an environment.
1)
28