5.12 Modification of the product by the user
Erbe Elektromedizin expressly warns against modifying the
product. Any modification exempts Erbe Elektromedizin from
any and all liability.
6
Unit settings
This chapter describes unit settings that may be used to operate
the product without causing damage.
You can find setting recommendations for various application ar-
eas in the Erbe brochure for gastroenterology.
When operating this product, you must observe the setting limits
specified in these instructions. Adapt the power to the planned
procedure. Begin with the lowest possible power setting and in-
crease gradually as required.
6.1
Permitted settings on the ERBEJET 2
Rinsing with to high of a pressure can lead to tissue damage.
Therefore always select for rinsing the lowest possible effect on
the ERBEJET 2.
Effect levels 1 – 80 of the ERBEJET 2 may be used. Please match
the effect and thus pressure of the water jet to the organ you are
working on.
6.2
Permitted settings on the Erbe VIO D series, VIO S
series and VIO 3
The instrument detection automatically sets suitable operating
parameters on the device (known as default values). You can
subsequently change these operating parameters, but only
within a certain range.
27