Alarm Output(S) - Siemens POINTEK CLS100 Instrucciones De Operación

Ocultar thumbs Ver también para POINTEK CLS100:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 67

Alarm Output(s)

• mA
• solid state switch
• relay output
- max. switching voltage
- max. switching current
- max. switching power
• intrinsically safe
• repeatability
WARNING: A wet location is a location where water or other
conductive liquid may be present and is likely to increase the risk of
electric shock.
Environmental
• ambient temperature:
- 316L SS process connection
(integral cable or enclosure
version)
- fully synthetic process
connection (enclosure version)
• ingress protection:
- Enclosure version
- Integral cable version
• installation category:
• pollution degree:
Mechanical
• common probe/wetted parts: PPS process connection and PPS sensor; or 316L
process connection and PPS or PVDF sensor.
[Standard Metal process connection seal is FKM (e.g. Viton). FFKM (e.g. Kalrez) is optional].
• cable version (SS process connection):
• integral cable body: 316L stainless steel
• 316L stainless steel process connection:
3/4" NPT or R 1" (BSPT), or G 1" (BSPP)
Page EN-2
Pointek CLS100 – QUICK START MANUAL
Fully synthetic
process connection,
enclosure version
4/20 mA or 20/4 mA 2-wire current loop detection
not available
60 V DC or 30 V AC;
limited to 30 V DC/16 V AC
in wet locations
1 A
60 W
not available
2 mm (0.08")
-30 to +85 °C (-22 to +185 °F)
-10 to +85 °C (+14 to +185 °F)
Type 4 / NEMA 4 / IP68
Type 4 / NEMA 4 / IP65
I
4
• 1 m (3.3 ft) of 4 conductor, 22 AWG,
shielded, polyester jacket
Stainless steel process con-
nection, integral cable or
enclosure version
30 V DC/30 V AC
82 mA max.
Limited to 30 V DC/16 V AC
82 mA max. in wet
locations
not available
30 V DC max. (suitable
barrier required)
A5E32146158

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Pointek cls100

Tabla de contenido