Installation

Do not install or use the pump in an environment exposed to
atmospheric agents (rain, snow, ice), dust, aggressive gases, or in
explosive environments or those with a high fire risk.
During operation, the following environmental conditions must be
respected:
Do not remove the adhesive and protective cap before connecting the turbopump
CAUTION!
to the system.
Figure 2
In the presence of magnetic fields the pump must be protected using
a ferromagnetic shield. See the appendix "Technical Information" for
detailed information.
The Turbo-V 1K-G Retrofit must be connected to a primary pump (see
"Technical Information").
Turbo-V 1K-G Retrofit User Manual / 87-900-996-01 (B)
maximum pressure: 2 bar above atmospheric pressure
temperature: from +5 °C to +35 °C (see the diagram pressure-
temperature in the appendix "Technical Information")
relative humidity: 0 – 95 % (non-condensing)
Instructions for Use

Installation