Switching on and Use of Turbo-V 1K-G Retrofit

The following instructions apply to the Turbo-V 1K-G Retrofit used
togheter with Navigator controller (model 9698978M004). Please
refer to the relative control unit manual for use with other controller
models (9699454M001 and 9699554M003).
To switch on the Turbo-V 1K-G Retrofit it is necessary to supply the
mains. The integrated controller automatically recognizes the
interlock and start signals presence and start up the pump.
The first pump start up is in "Soft Start" mode. When the start up
cycle is finished, the "Soft Start" mode automatically is disabled, and
the following start ups are without the "Soft Start" mode. To re-
enable the "Soft Start" mode it must be activated by the suitable
software command (see the paragraph "RS 232/485
COMMUNICATION DESCRIPTION" in the appendix "Technical
Information").
The green LED located on the Navigator controller base front panel
indicates with its flashing frequency the system operating conditions:
Turbo-V 1K-G Retrofit User Manual / 87-900-996-01 (B)
with no flashing: the pump is normally rotating;
slowly flashing (period of about 400 ms): the system is in ramp,
or in braking, or in Stop, or in "Waiting for interlock" status;
fast flashing (period of about 200 ms): error condition.
Instructions for Use
Use
235/294