INSTALLING THE OPTIONAL FR3 DATA CARD AND THE BATTERY
The FR3 automatically stores summary data of last use in internal memory
for later review on the display screen. However, if your protocol requires
recording detailed information about the operation and use of the FR3, you
must install an FR3 data card, available separately.
1.
2.
Install the battery.
Press ON for Patient Use
As soon as the battery is installed, the FR3 automatically runs an interactive user-initiated test, if it has
Clock Not Set
Set Date and Time
not already done so following removal of the language card, as described previously. Press the buttons
when directed, or the test will fail.
Battery
Good
Data Card
Good
Administration
OFF
SETTING THE DATE AND TIME
The first time you set up the FR3, it automatically runs a user-initiated test when the battery is
inserted. When the test is complete, the FR3 displays a reminder that the clock is not set and a
ADMINISTRATION
##
Incident Data
#:##:##
prompt to set the date and time.
User-Initiated Test
System Information
The date format of the FR3 is the year in four digits, the name of the month, and the day in two digits.
Setup
Wireless Data Transfer
For example: 2010 February 17. The time format uses a 24-hour clock and shows two digits each for
the hour, minute, and second. For example: 18:05:54. To check the clock and modify it, if needed, to
Setup
match local date and time, follow these steps:
Back
Load New Setup
1.
View Setup
Save Setup to Data Card
2.
Date and Time
3.
4.
* Non-Philips data cards may not work properly with the FR3 system.
† If the FR3 is already in standby, press the green On/Off button to turn on the FR3. When voice prompts begin,
press the button again to display the status screen.
Insert the optional data card into the data card slot in the bottom of the
FR3 battery compartment. A label on the floor of the battery
compartment shows the correct orientation for inserting the card. A
retention clip helps hold the card in the slot. The tail on the data card
serves as a visual reminder that a card is installed and aids in card
removal.
Check the battery label to be sure the battery is within its install-by date. Remove the battery
from its packaging. Install the battery in the FR3 battery compartment as shown. Be sure that both
the battery's latches click into place.
Press the ADMINISTRATION option button to go to the ADMINISTRATION screen.
On the ADMINISTRATION screen, press the option button to scroll to SETUP, then press the
option button to select it.
On the SETUP screen, press the option button to scroll to DATE AND TIME, then press the
option button to select it.
On the DATE AND TIME screen, press the option button to stop the automatic time
incrementing and display the CHANGE DATE AND TIME screen.
*
(Optional) Install the
FR3 data card.
†
7