Temperature-Related Messages
Message
Temp: Probe Failure
Temp: Probe
Disconnect
Temp: Internal Error
SpO
-Related Messages
2
SpO
: No Sensor
2
SpO
: No Si
2
SpO
: Motion Artifact
2
SpO
: Internal Failure
2
General Monitor Messages
SYSM: Battery Low
SYSM: Battery Empty
System: Internal
Failure
SYSM: Check Clock
SYSM: Battery Near
Empty
SYSM: Battery
Temperature
SYSM: Factory
Defaults
SYSM: Failure xxxx
Printer-Related Messages
PRNT: Check Printer
PRNT: Paper Empty
PRNT: Internal Error

Battery Care and Replacement

You can recharge the battery at any time by plugging the power cord into a properly grounded, 3-wire,
hospital grade AC power source. The battery indicator displays the charging state of the battery. Bringing
the charge from a completely discharged state to above 90% of capacity can take up to four hours. To
ensure maximum battery power, keep the monitor plugged in to an AC power source when stationary.
The battery can be accessed from the bottom of the unit and should be replaced by qualified personnel only.
Philips recommends that the battery be replaced every year to guarantee optimum battery life. Disposal of
the old battery should follow local and national recommendations.
Description
Defective temperature probe.
The temperature probe is disconnected
or defective.
Temperature module failure.
Loss of probe signal.
gnal
Patient perfusion too weak
Too much patient movement
SpO
module failure.
2
Battery is below 20% of its capacity
Battery is empty
Monitor failure
RTC potential problem
Battery is below 10% of its capacity
Internal temperature is greater than
o
o
F (
C)
122
50
System returned to factory default
settings.
ROM Check Sum error
The printer platen is up.
The printer paper is empty.
Printer failure.
Battery Care and Replacement
Action Required
Cycle the power on, then off. If the failure
continues, replace the probe.
Check connections and hose. Replace if
necessary.
Contact the Philips Response Center.
Verify attachment of sensor to extension cable
and to patient finger.
Check placement of sensor and if polish is on
finger. Remove polish and reapply sensor.
Check for patient movement and check
placement of sensor.
Contact the Philips Response Center.
Connect unit to suitable AC power outlet.
Connect unit to suitable AC power outlet.
Contact the Philips Response Center.
Reset time through System Setting screen.
Connect unit to suitable AC power outlet.
Cool down unit. Replace battery.
No action required if user selected the factory
defaults. If the monitor shows this indication
frequently, replace the main board.
Replace main board.
Lower the printer platen.
Insert a full roll of printer paper.
Remove any obstacles from the printer,
including paper jams. Reset the power.
Replace the main board or printer as required.
Troubleshooting and Maintenance