Instructions for Use
Figure 19 MS40-S packaging

Installation

Do not install or use the pump in an environment exposed to atmospheric agents
(rain, snow, ice), dust, aggressive gases, or in explosive environments or those with
a high fire risk.
During operation, the following environmental conditions must be respected:
Before starting the pump, fill up with lubricating oil as the pump is delivered
CAUTION
empty.
Oil must be poured into the casing through the special threaded plughole and
CAUTION
NOT through the suction line.
It is important that the oil level stays within the range MIN and MAX that is
NOTE
indicated by the level gauge on the side of the pump.
182
temperature: from +12 °C to +40 °C
relative humidity: 0 ÷ 95 % (non-condensing)
MS40-S User Manual