Instructions for Use
Table 7
Oil level checking
Oil change
Anti-Suckback valve checking
Fan cover cleaning
Exhaust filter replacement if oil mist at exhaust or yearly.
The frequency with which repairs are performed depends on the process and
NOTE
presence of substances that shorten pump life (dust, abrasives, solvents, water,
chemically aggressive substances).
Use only AVF Platinum oil. The use of oils other than the one foreseen alters the
pump performance, could damage the pump and in any case invalidate the
warranty.
190
SCHEDULED MAINTENANCE: Maintaining the nominal state of operation.
UNSCHEDULED MAINTENANCE: Restoring the nominal state of operation.
Daily (before every starting)
8.000 hours (light applications)
6 months
6 months
MS40-S User Manual