Caution Notes; Protective Equipment - Agilent Technologies MS40-S Manual Del Usuario

Tabla de contenido

Publicidad

Instructions for Use

Caution notes

Protective equipment

The protective equipment of the operators who are operating or executing the
maintenance of the pumping system must always be adequate for the type of
operation being executed. Furthermore, it must comply with the safety
requirements of the legislation in force in the country in which the machine is used.
In general, the operator must wear accident-prevention shoes while handling the
MS40-S and during installation.
MS40-S User Manual
Before putting the pump back into operation after a breakdown, inspect it and
check carefully for any other signs of damage.
Use only tools that are in perfect working order and specially designed for the
job; use of inappropriate or ineffective tools may cause serious damage.
Always check the lubricant and that it is properly distributed through the pump;
inadequate lubrication may damage the pump seriously.
Give the parts some form of marking as you strip them down to ensure that
you reassemble them again in the proper order.
Check that there are no scratches or grooves on the machined shafts, in their
seats inside the pump or on machine-ground surfaces. Slight scratches and
abrasions may be eliminated with very fine emery paper or by a little light
grinding.
Before putting a group together, always spread a little oil over inner parts and
mating surfaces. Replace all seals with original spare parts before
reassembling components.