Technical Information
Window Meanings
Table 16 Window meanings
WIN
000
008
1098
120
127
202
203
205
206
216
223
233
300
301
302
307
319
320
321
322
323
358
362
214
TYPE
R/W
DESCRIPTION
N
R
START/STOP (1= START; 0= STOP)
N
R
REMOTE/SERIAL CONFIGURATION (1= Remote; 0= Serial)
N
R
SERIAL BAUD RATE (0-5) [0=600, 1=1200, 2=2400, 3=4200, 4=9600,
5=19200 bit/s]
N
R
ROTATIONAL FREQUENCY SETTING [Hz] (40 - 60)
N
R
ROTATIONAL FREQUENCY SETTING [rpm] (read only)
N
R
ELECTRICAL POWER DELIVERED TO THE MOTOR [W]
N
R
ELECTRICAL DRIVING FREQUENCY [Hz]
N
R
PUMP STATUS
(0=stop; 1=wait interlock; 2=start; 3=autotuning; 5=normal; 6=fail)
N
R
ERROR CODE:
(Bit 7: Motor blocked; Bit 6: shortcircuit; Bit 5: Undervoltage/Overvoltage;
Bit 4: Motor overtemp; Bit 3: Verify line; Bit 2: Controller overtemp; Bit 1:
Bus Undervoltage; Bit 0: overcurrent)
N
R
CONTROLLER TEMPERATURE [°C]
N
R
PUMP BODY TEMPERATURE [°C]
N
R
ROTATIONAL FREQUENCY [rpm]
N
R
CYCLE TIME [min]
Time elapsed since start
N
R
CYCLE NUMBER
N
R
PUMP LIFE [h]
N
R
PUMP LIFE IN STAND BY (STOP STATUS) [h]
A
R
CONTROLLER MODEL NUMBER
A
R
PUMP MODEL NUMBER
A
R
PUMP SPECIAL MODEL NUMBER
A
R
PUMP SERIAL NUMBER
A
R
CONTROLLER SERIAL NUMBER
N
RW
MAINTENACE TIMER [hours]
N
R
LAST WORKING HOUR TEMPERATURE AVERAGE [°C]
MS40-S User Manual