Instructions for Use
It is mandatory to leave ample free space all around the pump in order to allow
CAUTION
proper air circulation; inserting the pump in a closed volume without air
circulation is not allowed. For an internal safety system, the pump exceeded the
ambient temperature of 40 °C, could show an error.
Keep the protective caps placed on the inlet and exhaust connections and
WARNING
do not switch on the pump until they are properly connected to the
instrument and the exhaust line respectively.
During installation, pay maximum attention that the suction flange is
WARNING
connected to the vacuum chamber and the exhaust connection is not
closed (see the following figure). The pump must not be used as a
compressor.
Non-observance of these precautions may be dangerous for the machine
and the operator.
Inlet flange
Figure 20 Flanges and connection
MS40-S User Manual
Exhaust
connection
183