Analogue calibre
Resolution
Accuracy
Digital
calibres
0.19 MΩ
Resolution
Accuracy
3% R ± 5 ct
Test voltage
Voltage at
open circuit
Test current
Short-circuit
current
Once the yellow key has been released, the residual voltage
present on the terminals is discharged via the measuring cords
through the apparatus at the speed of 1.5 s/µF and in an
impedance of 300 Ω.
Average charging time depending on the capacity in
parallel on the resistance to be measured.
1 MΩ
100 MΩ
6.2.3. Continuity
Measurement range : 0 to 20 Ω
Calibre
Resolution
Accuracy
Measuring current
Voltage at open circuit
6.2.4. Resistance (C.A 6523 and C.A 6525)
Measurement range : 0 to 400 kΩ
Calibres
0.0...399.9 Ω 400..3999 Ω 4.00..39.99 kΩ 40.0..399.9 kΩ
auto.
Resolution
Accuracy
± 3% L ± 5 pt
Test
current
Voltage at
open circuit
8 segments per
5 % L ± 1 segment
0.01 to
0.20 to
39.99 MΩ 399.9 MΩ
10 kΩ
250 V
< 300 V
≥ 1 mA
for R ≤ 250 kΩ
500 V
3 s
3 s
6 s
20 s
0.00...19.90 Ω
10 mΩ
± 3% L ± 1 pt
≥ 200 mA
7 V ≤ V
0.1 Ω
55 or 550 µA according to the measurement
7 V ≤ V open ≤ 9 V
50 kΩ...2 GΩ
10-unit interval
40.0 to
100 kΩ
3% R ± 2 ct
500 V
< 600 V
≥ 1 mA
for R ≤ 500 kΩ
≤ 3 mA
1000 V
3 s
6 s
6 s
20 s
≤ 9 V
open
1 Ω
10 Ω
± 3% L ± 1 pt
37
400 MΩ
to 2 GΩ
1 MΩ
1000 V
< 1200 V
≥ 1 mA
for R ≤ 1 MΩ
1 µF
5 µF
1 µF
5 µF
100 Ω