clottrac_chap.fm 10/27/04 10:55 am
3.875 x 3.187 inches (98 mm x 81 mm)
c.
Add one drop (40 µl) of 0.1 M CaCl
immediately close the actuator heat block to initiate the test.
7.2 RACT Cartridge:
a.
Transfer 0.2 mL of reconsituted control into each channel of the cartridge.
Fill to between the fill lines etched of each channel body.
b.
Place the filled cartridge into the instrument. Ensure the incubate switch is
on.
c.
Close the actuator heat block to initiate the test. The test will automatically
initate after 300 seconds (5 minutes).
7.3 GPC Cartridge:
a.
Transfer 0.1 mL of aPPT, PTT, or thromboplastin reagent into each channel
of the cartridge, depending on the test being performed.
b.
Transfer 0.1 mL of reconstituted control into each channel of the cartridge.
Refer to the reagent package insert for specific instructions and incubation
times.
c.
If the specified incubation time for the reagent is 5 minutes, ensure the
incubate switch is on. Close the actuator heat block the test will
automatically initate after 300 seconds (5 minutes).
8.
For interpretation of data, refer to the Results section.
A08718002-01
Medtronic Confidential
CS003
to each channel of the cartridge and
2
English
Instructions For Use 7