Motor stall. Power OFF then ON (E9).
System error. Power OFF then ON.
Bad Reference in Ch1 and Ch2.
OR
Bad Reference in Ch1.
OR
Bad Reference in Ch2.
Error: E1
Error: E2 or E3
No flags.
OR
No flag in Ch1.
OR
No flag in Ch2.
OR
No flags in Ch1 or Ch2.
Insufficient data.
Error: Invalid reply.
Light path failure in Ch1 and Ch2 (E4).
OR
Light path failure in Ch1 (E4).
OR
Light path failure in Ch2 (E4).
Motor stall.
Motor stall, cycle power (E8).
No reply.
File system error.
Comm. Error; Invalid data.
Notes: If a System Alarm occurs that cannot be resolved by cycling the power OFF
then ON, record the 3-digit Error Code (if displayed), and contact Medtronic
Instrument Service for assistance.
48 English
86702-001
Rev 1.0
The motor has been inhibited. To reset the
system, turn the power OFF then ON.
The system has become unstable. To reset
the system, turn the power OFF then ON.
The ACT Plus™ instrument could not get a
baseline drop time for Channel 1, Channel 2,
and both channels. Repeat test.
No cartridge is detected. Place a cartridge in
the actuator assembly and retest.
A software error has occurred. Call
Medtronic Instrument Service.
The ACT Plus™ instrument did not sense a
flag in Channel 1, Channel 2, or both
channels. Repeat the test.
The channels of the control timed out at 999.
Repeat the control.
Internal communication error. Call Medtronic
Instrument Service.
The flag sensor has failed. Clean the
actuator assembly cover per the instructions
in the Maintenance section of this manual. If
the error is not corrected, call Medtronic
Instrument Service.
The motor is inhibited. Press STOP to reset.
Repeat the test.
The motor is inhibited. To reset, power OFF
then ON.
Internal communication error. Call Medtronic
Instrument Service.
Call Medtronic Instrument Service.
Call Medtronic Instrument Service.