Scanning In Hand-Held Mode; Setting Bar Code Format Options; Code 39 - Medtronic ACT Plus Manual Del Usuario

Ocultar thumbs Ver también para ACT Plus:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

ACT_A.fm 6/23/04 1:23 pm
UC200304151c ML
5.5" x 7.75" (139.7 mm x 196.85 mm)

Scanning in Hand-Held Mode

1. Aim the scanner at the bar code and press the trigger.
2. Be sure that the red line from the scanner goes across the entire bar code.
When the scanner has successfully decoded the bar code, the scanner will beep,
and the information will be entered into the ACT Plus™ instrument.

Setting Bar Code Format Options

The scanner is preset to read UIDs and PIDs in format Code 39 With Check Digit
(Mod 10), Interleaved 2 of 5 with Check Digit (Mod 49), Codabar and Code 128. It
is also preset to read the cartridge/ control information.
To set additional barcode format options:

Code 39

Scan the "Enable Code 39 Check Digit" if the UID/PID barcodes use check digit. To
enable the check digit in Code 39 to verify a correct scan.
Scan the "Disable Code 39 Check Digit" figure if the UID/PID barcodes do not use
check digit verification. Disable the check digit in Code 39 to verify a correct scan.
Note: Disabling the Code 39 Check Digit will still allow the scanner to scan a
barcode that containsthe check digit and add it to the data.
52 English
86702-001
Rev 1.0
Figure 9. Correct scan location
Figure 10. Enable Code 39 Check Digit
Figure 11. Disable Code 39 Check Digit
Medtronic Confidential
ACT Plus™ Operator's Manual
CS010

Publicidad

Tabla de contenido
loading

Tabla de contenido