To Change The Lot Number Of The Currently Active Cartridge; To Remove A Cartridge Lot; Enter A Cartridge Lot Number And Expiration Date Using The Optional Bar Code Scanner; View Current Test Results For The Last Test Performed - Medtronic ACT Plus Manual Del Usuario

Ocultar thumbs Ver también para ACT Plus:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

ACT_CH.fm 6/23/04 1:23 pm
UC200304151c ML
5.5" x 7.75" (139.7 mm x 196.85 mm)
When there are two lot numbers entered for a cartridge type, the following variable
function keys will appear: [Toggle Active], [Remove Lot], [Edit Lot/Exp Date] and
[Exit to Main Menu].

To change the lot number of the currently active cartridge:

1. Select [Toggle Active].
*
2. Move the "
" to the cartridge lot number that is currently in use.

To remove a cartridge lot:

1. Select [Remove Lot].
2. To select the lot to be removed, use the up/down arrows (↑↓) to toggle to the
appropriate lot number.
3. To delete the selected cartridge lot number, select [Remove Selected Lot].
Enter a Cartridge Lot Number and Expiration Date Using
the Optional Bar Code Scanner
From the Main Menu:
1. Select [Cartridge Lot]. Page 1 of the Cartridge Lot/Expiration Date screen
will appear.
When there is no or one lot number for a cartridge type:
1. Scan the bar code on the cartridge box. The lot number and expiration date
will automatically populate their respective fields.
2. To return to the Main Menu, press [Main Menu].
When two lot numbers exist for a cartridge type:
1. Select [Remove Lot].
2. To remove one of the lot numbers, toggle to the appropriate lot number.
3. To delete the selected cartridge lot number, select [Remove Selected Lot].
4. Scan the bar code on the cartridge box. The lot number and expiration date
will automatically populate their respective fields.
5. To return to the Main Menu, select [Main Menu].

View Current Test Results for the Last Test Performed

1. To display the results of the last test, select [View Current Test Results].

Transmit Patient and Quality Control (QC) Test Results

Patient and QC test results can be transmitted to a floppy drive or serial port. The
transmit destination is set from the Instrument Parameters Menu.
If transmitting to a floppy drive, insert a floppy disk before continuing.
22 English
86702-001
Rev 1.0
Medtronic Confidential
ACT Plus™ Operator's Manual
CS010

Publicidad

Tabla de contenido
loading

Tabla de contenido