Figure 9 Carefully pull the back casing upwards and slide it to the right slightly
Carefully open the back casing, while the internal FPC cable is still connected.
Figure 10 Open the back casing carefully ane remove adhesive tape were nesessary
i)
Remove adhesive tape
If adhesive tape is protecting the connectors on the boards, you can remove it (see Figure 10). It is not necessary to
replace the tape, as it is only for protection during transportation to the end user.
j)
Release the FPC cable
Flip the white lock upwards with a fingernail (see Figure 11).
2 cm
Figure 11 Disconnect the FPC cable
1 cm
13