6.
Close the jaws to grasp the tissue. Do not grasp too much
tissue.
The tissue is grasped.
7.
Lock the handle.
The tissue is clamped.
8.
Activate the HF current for coagulation using the finger
switch on the handle or footswitch:
9.
Release the finger switch or footswitch.
10. Press the handle further down to release the lock and to
open the jaws.
The tissue is coagulated.
Cutting the tissue
The tissue is grasped by the jaws and is coagulated.
WARNING
Strong bleeding may occur if the grasped tissue is cut
before coagulation or ligation!
1.
To cut, pull the trigger and then release it.
The tissue is cut in the centre of the sealed area.
MN031-600-S0 EN
A continuous acoustic signal sounds during the
entire sealing process to indicate that power is
being supplied.
An alternating acoustic signal indicates the end
of the sealing process.
If there is a short-circuit in the area of the
sealing instrument or if the sealing instrument
is not in contact with tissue during activation, a
warning will be displayed on the HF device.
Please check the instrument and reactivate.
Before cutting, ensure that the tissue has been reliably
sealed.
Cut only in the sealed area.
ERGO 310D
11