Commissioning

type of sensor
setting ?
Access to all setting menus can be blocked with an access code !
meas. parameter
setting ?
cali. zero point
PAR 1
34
8
Commissioning
8.1

Presettings

Check first that the sensor is correctly connected to the
DULCOMETER
®
D1C (do not open the sensor).
type of sensor :
ICT 1
sensor data
are changed!
Check that the right sensor type has been selected (see "type of
sensor setting?")
mr: meas. range
cc: cellconstant
α: temperature-
coefficient
Set the measurement range of the sensor (see "meas. parameter
setting?". Bypass the second parameter set (press Enter))
IMPORTANT
Check settings in all menus when changing the range allocation.
8.2

Zero point calibration

cali. zero point
probe in air!
accept value?
12 µS/cm
IMPORTANT
• The zero point must be calibrated during commissioning.
• The zero point must be calibrated if the measurement range
is changed.
20 mS/cm
mr :
cc :
1.120/cm
α :
2.10% °C
15 mS/cm 23.8 °C
please wait!
= automatic timing
mr 2 : 200 µS/cm
cc 2 : 1.120 /cm
α:
2.10% /°C
1106 µS/cm 25.0 °C