Foot Switch Settings; Handles With Buttons; Neutral Electrode Settings And Attachment - HEBU medical HBS x-touch 300 argon BiSeal Instrucciones Para El Uso

Aparato de cirugía electrodoméstica
Ocultar thumbs Ver también para HBS x-touch 300 argon BiSeal:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

EN

6.5 Foot switch settings

After connecting the foot switch at the back, the required mode must be set at the touchscreen. To set
the mode, touch the field indicating the foot switch symbol. This changes the text which appears
above the symbol.
It is important to ensure that the instruments are connected to the connectors, which are selected by
footswitch. For example: In Monopolar 1 connected instruments can only be operated with the
settings MONO1/MONO2 or MONO1 + BIAUTO of the foot switch.

6.6 Handles with buttons

Activating the handle with buttons takes place independently of the foot switch settings. These cannot
be activated using the foot switch.

6.7 Neutral electrode settings and attachment

In the monopolar operating mode, a neutral electrode is required.
After attaching the neutral electrode to the patient, the HF unit recognize the neutral
electrode automatically (cf. illustration on the left and chapter 3.2 "Overview of symbols
for operating modes and settings").The HBS x-touch 300 argon BiSeal
neutral electrodes with divided and undivided contact surfaces.
!
If the displayed neutral electrode type does not agree with the used
neutral electrode on the patient, a defective neutral electrode or a
Please note the following information when attaching the
46
neutral electrode type, this is indicative of poor adhesion of the
defective cable!
neutral electrode
HBS x-touch 300 argon BiSeal
®
recognizes
®

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Hf 9516

Tabla de contenido