Saving And Accessing The Program - HEBU medical HBS x-touch 300 argon BiSeal Instrucciones Para El Uso

Aparato de cirugía electrodoméstica
Ocultar thumbs Ver también para HBS x-touch 300 argon BiSeal:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

EN
Argon gas flow setting
The gas flow can be set at an interval of 0.1 L/min to 6.0 L/min in steps of 0.1 L/min.
Different gas flows can be set for each of the functions CUT and COAG.
To set the gas flow, touch the relevant display - CUT or
COAG. This will open the menu for setting the gas flow. Use
the arrow keys to set the required output.
To save and return to the main menu, select the "Exit" button.

6.9 Saving and accessing the program

Touching the "Program" field causes a selection of saved
programs to appear.
To select the required program touch the relevant line. The
program you are located in is highlighted in grey.
In the main menu, the number of the accessed program
and the name are displayed
To change the program name, touch the field with the name.
In our example, this is: DR MULLER. A window opens with a keypad. The maximum symbol length
is 15 characters.
To save the name, press "Enter", and to quit without saving, press "Esc".
Every change of the settings is automatically saved in the
program you are currently in.
In order to save individual settings in a new program,
initially select a free storage location. Settings and the
name of the program can then be changed at will.
50
HBS x-touch 300 argon BiSeal
DR. MULLER
1
®

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Hf 9516

Tabla de contenido