Oscillo-Saw GOLD II Autopsy
3.2 Inserting the new saw blade
1. Check that the female thread and saw blade support of the drive shaft are clean.
2. Position
It makes no difference which side of the saw blade is facing inwards or outwards.
Important!
The notches in the saw blade must be positioned precisely flush on the spigots of the
drive shaft.
3. Position the fixing ring. The notches in the fixing ring must always be pointing towards
the saw (see fig. 2). These must also be positioned precisely on the spigots.
4. Screw the retaining screw in manually until hand tight.
5. Then
tighten
The tightening torque is 6 - 7 Nm. Caution! Do not overtighten.

4 Operation

Switching on and off (see fig. 3)
Switching on:
Push the switch forwards
Switching off:
Push the switch back
Setting the oscillating frequency (see fig. 4)
Select the oscillating frequency when the motor is running.
The optimum oscillating frequency can be selected at the s
peed gauge depending on the properties of the plaster cast.
Oscillating frequency preselection using the electronic
speed gauge.
Lowest oscillating frequency
Highest oscillating frequency
the
new
firmly
using
the
Inserting the saw blade
( I )
( 0 )
( 1 )
( 8 )
saw
fork
wrench,
as
Fig. 2:
O
EN
blade.
described
above.
I
Fig. 3
Fig. 4
19