Ivoclar Vivadent Bluephase Instrucciones De Uso página 87

Ocultar thumbs Ver también para Bluephase:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 58
Guidance and manufacturer's declaration - electromagnetic immunity
The "Bluephase (G2)" is intended for use in the electromagnetic environment specified below. The customer or the user of the "Bluephase (G2)"
should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Conducted RF
3 V
rms
IEC 61000-4-6
150 kHz to 80 MHz
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to 2.5 GHz
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
a
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV
broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
Recommended separation distances between portable and mobile RF communications
equipment and the "Bluephase (G2)"
The "Bluephase (G2)" is intended for use in the electromagnetic environment in which radiated RF disturbances are controlled.
The customer or the user of the "Bluephase (G2)" can help prevent electromagnetic interference by maintaining a miminum distance between portable
and mobile RF communications equipment (transmitters) and the „Bluephase (G2)" as recommended below,
according to the maximum output power of the communication equipment.
Separation distance according to
Rated
maximum
output power
of transmitter
150 kHz to
W
80 MHz
d = 0.4 P
0.01
0.04
0.1
0.13
0.40
1
1.3
10
4.0
100
For transmitters rated at a maximum output power not listed above, the recommended
separation distance d in metres (m) can be determined using the equation applicable to
the frequency of the transmitter, where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency
range applies.
NOTE 4: These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
Compliance level
10 V
10 V/m
frequency of transmitter
m
80 MHz to
800 MHz to
800 MHz
2.5 GHz
d = 0.4 P
d = 0.7 P
0.04
0.07
0.13
0.22
0.40
0.70
1.3
4.0
Electromagnetic environment - guidance
Portable and mobile RF communications equipment should not be used closer to any part
of the „Bluephase (G2)", including cables, than the recommended separation distance
caculated from the equation applicable to the frequency of the transmitter.
Recommended separation distance
d = 0.35 P
d = 0.35 P 80 MHz to 800 MHz
d = 0.70P 800 MHz to 2.5 GHz
where P is the maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer and d is the recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an electromagnetic site
survey,
should be less than the compliance level in each frequency range.
a
Interference may occur in the vicinity of equipment marked with the following symbol:
environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which the „Bluephase (G2)"
is used exceeds the applicable RF compliance level above, the „Bluephase (G2)" should
be observed to verify normal operation. If abnormal performance is observed, additional
measures may be necessary, such as reorienting or relocating the „Bluephase (G2)".
Over the frequency range 150 kHz to 80 MHz, field strength should be less than 10 V/m.
b
2.2
7.0
b
87

Publicidad

Tabla de contenido
loading

Tabla de contenido