Connect the power cord with the power supply and the
power pack. Make sure that the power cord is easily
accessible at all times and and can easily be disconnect-
ed from the power supply. The charging base will briefly
display "Bluephase G4" as ticker text on the screen.
Handpiece
Unpack the handpiece from the packaging and detach
the light guide by pulling it out. Then clean the
handpiece and light guide (see Maintenance and
Cleaning). After cleaning, reinsert the light guide.
For reasons of hygiene, we recommend using a
disposable protective sleeve for each patient
(see Maintenance and Cleaning). Make sure to fit the
protective sleeve snugly to the light guide. Sleeves
enclosed in the delivery form may be used or sleeves can
be purchased per country specific regulations.
Then, attach the anti-glare cone or anti-glare shield to
the light guide.
Battery
The battery must be fully charged before it is used for
the first time! When fully charged, the battery features a
curing capacity of at least 20 minutes.
Slide the battery straight into the handpiece until you
hear and feel it click into place.
Gently place the handpiece in the corresponding rest in
the charging base without using any force. If a hygiene
sleeve is used, please remove it before you charge the
battery. If possible, always use the light with a fully
charged battery. This will prolong the service life. We
therefore recommend placing the handpiece into the
charging base after each patient. If the battery is
fully discharged, the charging time is 2 hours.
The battery is an expendable part that typically needs to
be replaced after approximately 2.5 years. See battery
label for the age of the battery.
Battery: charging status
The current charging status is displayed on the
handpiece as described on page 6.
11