Electromagnetic compatibility (EMC, IEC 60601-1-2:2014 4th Edition,
Table 6)
Recommended separation distance between portable and mobile RF communications
equipment and the Invia Motion pump
The Invia Motion pump is intended for use in an electromagnetic environment in which
radiated RF disturbances are controlled. The customer or the user of the Invia Motion pump can help
prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the Invia Motion pump as recommended below, according to
the maximum output power of the communications equipment.
Rated maximum out-
put power
of transmitter
W
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommened separation distance (d)
in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is
the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
34
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
P
d = 0.35
0.035
0.11
0.35
1.11
3.5
m
80 MHz to 800 MHz
P
d = 0.35
0.035
0.11
0.35
1.11
3.5
800 MHz to 2.5 GHz
P
d = 0.7
0.07
0.22
0.7
2.21
7.0

Publicidad