DJO ARTROMOT-K1 Instrucciones De Uso página 49

Ocultar thumbs Ver también para ARTROMOT-K1:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 104
− When disconnecting the device from
the power line, remove the plug from
the wall outlet first, before disconnec
ting the cable from the device.
− When connecting the device to other
equipment or when creating a medi
cal system, check that the sum of
leakage currents will not cause any
hazard. Please contact DJO, if you
have questions in this matter.
− Do not use multiple portable socket
outlets MPSO to connect the device
to the power line. The ARTROMOT® K
must be connected to a properly in
stalled wall outlet with a non fused
earthed wire.
Before connecting the power cord,
it must be completely unrolled and
placed such that it will not get caught
in the moving parts of the device.
− Before cleaning and service interven
tions, disconnect the device from
the power line by removing the
power cord from the wall outlet.
− Liquids must not be allowed to enter
the CPM device or the control pen
dant. If liquids have entered into the
devices, the ARTROMOT® K must
be immediately checked by a service
technician, before it can be reused.
Warning!
Equipment malfunction
− Magnetic and electrical fields are
capable of interfering with the proper
performance of the unit. For this
reason make sure that all external
devices operated in the vicinity of the
device comply with the relevant EMC
requirements. X ray equipment, MRI
devices, radio systems, cell phones,
etc. are possible sources of interfer
ence as they may emit higher levels of
electromagnetic radiation.
Keep the device away from such
equipment and verify its performance
before use.
− Refer repair and maintenance to
authorized persons.
Persons are authorized after training
by a specialist trained and commis
sioned by the manufacturer.
− Route all cables below the device
frame to either side, ensuring that
they cannot get caught by the moving
parts during operation.
− Check the ARTROMOT® K for dam
age and loose connections at least
once a year.
Damaged and worn parts must im
mediately be replaced with original
spare parts by authorized staff.
Caution!
Preventing chafing and pressure sores
When your patient is adipose, very tall
or very short, be sure to prevent chafing
and pressure sores. Place the leg con
cerned in a moderate abductive position
if deemed appropriate.
Caution!
Equipment damage
Check that the voltage and frequency
ratings of your local power line are
those indicated on the nameplate.
− The leg support element withstands a
maximum continuous load of
S <
,
/
kg S >
Do not allow any objects such as
blankets, cushions, or cables to get
caught in the moving parts of the
CPM device.
− Do not expose the ARTROMOT® K
to direct sunlight, because some of the
components may warm up excessively.
Furthermore, the device must be set
up at a safe distance from radiators to
prevent excessive temperature rises.
− The presence of children, pets and
rodents does not normally impair the
functioning of the device. However,
avoid contamination of the device by
children or animals, from dust and
lint, and keep them at a safe distance
from the device. The safety state
ments set forth apply.
kg
,
.

Publicidad

Tabla de contenido
loading

Tabla de contenido