Preparations At The Place Of Use; Preparation Before Cleaning; Cleaning/Disinfection; Product-Specific Safety Instructions For The Reprocessing - Aesculap Acculan 4 GA344 Instrucciones De Manejo

Ocultar thumbs Ver también para Acculan 4 GA344:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 46
en
7.3

Preparations at the place of use

Remove all attached components from the product (tool an accesso-
ries).
Remove any visible surgical residues as much as possible with a damp,
lint-free cloth.
Place the dry product in a sealed waste container and forward it on for
cleaning and disinfection within 6 hours.
7.4

Preparation before cleaning

Before the first automatic cleaning/disinfection: attach ECCOS holders
in a suitable tray.
Insert the products in the correct position into the ECCOS holders, see
Fig. D.
7.5

Cleaning/disinfection

7.5.1
Product-specific safety instructions for the reprocessing pro-
cedure
CAUTION
Damage to the product due to inappropriate cleaning/disinfecting
agents and/or excessive temperatures!
Use cleaning and disinfecting agents according to the manufac-
turer's instructions which
– are approved for plastic material and high-grade steel,
– do not attack softeners (e.g. in silicone).
Do not use cleaning agents that contain acetone.
Observe specifications regarding concentration, temperature and
exposure time.
Do not exceed the maximum temperature of 60 °C during chemical
cleaning and/or disinfection.
Do not exceed the maximum temperature of 96 °C during thermal
disinfection with FD water.
Dry the product for at least 10 minutes at a maximum of 120 °C.
CAUTION
Damage or destruction of the batteries due to processing!
Protect battery from moisture.
Note
The indicated drying time is a guide time only. It must be checked taking
into account the specific conditions (e.g. load) and if applicable adjusted.
8

Publicidad

Tabla de contenido
loading

Tabla de contenido