Note
The battery removal device can be used for easier battery removal, see
Fig. B.
►
Turn the product 1 with the battery shaft 4 upwards.
►
Press both safety catches 9 on the cap 8 at the same time and remove
the cap 8.
►
Tightly grasp the battery shaft 4 at the lower end.
►
Tap the battery shaft 4 on the palm of the hand until the battery 7
slides out of the battery shaft 4 and can be removed.
6.1.5
Protection against inadvertent activation
To prevent the product being inadvertently activated when changing a tool
or accessory, the knob for speed control can be locked.
Locking the knob for speed control 2:
►
Twist the safety catch 3 to position OFF.
The speed control knob 2 is blocked and the product 1 cannot be oper-
ated.
Unlocking the knob for speed control 2:
►
Twist the safety catch 3 to position ON.
The speed control knob 2 is unlocked and the product 1 can be oper-
ated.
6.1.6
Attaching and removing product attachments
Note
For additional information on the attachments for GA344/GA844, see
TA014552 and TA014553 (leaflet).
WARNING
Risk of injury when attaching/removing attachments/tools in the ON
position through inadvertent activation of the product!
►
Only attach/remove attachments/tools in the OFF position.
►
Secure product 1 from inadvertent activation with the safety catch 3,
see Protection against inadvertent activation.
Coupling
►
Push attachment 16 onto the product 1 until it clicks into place.
►
Pull on the attachment 16 to ensure a secure fit.
Uncoupling
►
Turn the sleeve 13 in the direction of the arrow 12 and simultaneously
remove the attachment 16 from the product 1.
Attach the Kirschner wire protection sleeve
Note
The special Kirschner wire chuck is recommended for the placement of
guide wires.
WARNING
Risk of injury when using long guide wires!
►
Use Kirschner wire protection sleeve when using long guide wires.
►
Screw on the Kirschner wire protection sleeve 15 into attachment 14.
Attach the tool to the attachment
Note
Attachment of the tools is described in operating instructions TA014552
and TA014553 (leaflet).
6
6.2
Function checks
The function test must be performed before each use and after each intra-
operative change of battery and attachment.
►
Make certain there is a battery in the battery compartment.
►
Ensure that the cap has completely clicked into place.
►
Check that the attachment is coupled properly. To test this, pull on the
attachment.
►
Check that the tool is coupled properly. To do this, pull on the tool.
►
For GB891R: check that the tool attachment has locked into place. To
do this, turn the tool attachment.
►
Ensure that the tool blades are not mechanically damaged.
►
Release product for use (ON position).
►
Run the product shortly at maximum speed in right and left rotation.
►
Make certain that the rotational direction is correct in each case.
►
Check the product for damage, abnormal operation noises, heavy vibra-
tion and excessive heat.
►
Do not use the product if it is damaged or defective.
►
Set aside the product if it is damaged.
6.3
Safe operation
WARNING
Coagulation of patient tissue or risk of burns for patients and user
through hot product!
►
Do not use products for acetabulum milling.
►
Cool the tool during operation.
►
Keep product/tool out of the reach of patients.
►
Let the product/tool cool down.
►
Use a cloth to protect against burns when changing the tool.
WARNING
Risk of infection from aerosol formation!
Risk of injuries caused by particles coming loose from the tool!
►
Use suitable protection (such as waterproof protective clothing,
face mask, safety gases, suction).
WARNING
Risk of injury and/or malfunction!
►
Always carry out a function check prior to using the product.
WARNING
Risk of injury when using the product beyond the field of view!
►
Apply the product only under visual control.
WARNING
Risk of injury and damage to the tool/system!
The rotating tool may get caught in drapes (such as textiles).
►
Do not let the tool come into contact with drapes (such as textiles)
during operation.
Note
The product drive motor is operated using a magnetic sensor system. In
order to prevent inadvertent activation of the motor, the product may not
be exposed to any magnetic fields (such as magnetic instrument pads).
Note
The slight whistling noise when the product is starting up is due to the con-
struction.