en
8.
Maintenance
To ensure reliable operation, the product must be maintained in accor-
dance with the maintenance labeling or at least once a year.
YYYY-MM
For technical service, please contact your national B. Braun/Aesculap
agency, see Technical Service.
9.
Troubleshooting list
►
Have defective products repaired by Aesculap Technical Service, see Technical Service.
Issue
Cause
Product not running
No battery
Battery not charged
Battery defective
Product in locked OFF position Safety catch is in OFF position
Product defective
Product becomes too hot Excessive use
Reprocessing/maintenance
carried out incorrectly
Fall damage, product defec-
tive
Use of a defective attachment Product heating
Blunt tool
Insufficient power
Product defective
Product is operating in left
rotation mode
Use of a defective attachment Attachment very hot
Blunt tool
Loud running noise
Defective product drive/ball
bearings
Defective attachment
drive/ball bearings
12
Detection
No battery in the battery shaft
No signal tone when inserting the bat-
tery
No signal tone when inserting the bat-
tery
Product not running
Product heating
Product heating
Product heating
Heating of tool and product
Insufficient product performance
Large amount of heat generated after a
short time
Serrated tool runs in counterclockwise
mode
Tool blades worn out
Loud, perceptible noise during opera-
tion
Loud, perceptible noise during opera-
tion
Remedy
Insert battery.
Charge battery in charger.
Have the manufacturer repair the battery.
Turn the safety catch into the ON position.
Have the manufacturer repair the product.
Follow operating instructions (operating mode).
Follow operating instructions (preparation, care).
Preventative: lubricate product before every steril-
ization.
Have the manufacturer repair the product.
Have the manufacturer repair the attachment.
Change tool.
Follow operating instructions (preparation, care).
Preventative: lubricate product before every steril-
ization.
Have the manufacturer repair the product.
Follow operating instructions (operating mode).
Have the manufacturer repair the product.
Operate serrated tool in right rotation.
Have the manufacturer repair the attachment.
Change tool.
Follow operating instructions (preparation, care).
Preventative: lubricate product before every steril-
ization.
Have the manufacturer repair the product.
Follow operating instructions (preparation, care).
Preventative: lubricate attachment before every
sterilization.
Have the attachment repaired by the manufac-
turer.