7.7.2

Mechanical alkaline cleaning and thermal disinfection

Machine type: single-chamber cleaning/disinfection device without ultrasound
Phase
Step
I
Pre-rinse
II
Cleaning
III
Intermediate rinse
IV
Thermal disinfection
V
Drying
D-W:
Drinking water
VE–W:
Vollentsalztes Wasser (demineralisiert, mikrobiologisch mindestens Trinkwasserqualität)
*Recommended: BBraun Helimatic Cleaner alkaline
Insert product 1 in the correct position in the ECCOS holder, see Fig. D.
Attach the rinsing adapter 17 (GA344211) to the product 1 in the
mount 14.
Then connect the interior rinse device and connect to the rinse connec-
tor of the cleaning/disinfection device/rinse cart.
Attach the Kirschner wire protection sleeve 15 to a rinse hose.
After automatic cleaning/disinfection, check visible surfaces for resi-
dues and repeat the cleaning/disinfection process as needed.
7.8

Inspection, maintenance and checks

Allow the product to cool down to room temperature.
Spray product after every cleaning and disinfection with oil spray
adapter 18 GB600880 (green) approx. 2 sec with Aesculap-STERILIT-
Power-Systems oil spray GB600, see Fig. C.
Note
Aesculap additionally recommends occasional spraying of moving parts
(such as knobs, couplings, cover lids) with the Aesculap STERILIT-Power-
Systems oil spray.
Check the product after each cleaning and disinfection for the follow-
ing: cleanliness, damage, function, abnormal operation noise, excessive
heat or heavy vibration.
Set aside the product if it is damaged.
7.9

Packaging

Always follow the instructions for use of the respective packaging and
storage devices (e.g. instructions for use TA009721 for Aesculap-
ECCOS holder system).
Insert the products in the correct position into the ECCOS holders, see
Fig. D.
Pack trays appropriately for the sterilization process (e.g. in Aesculap
sterile containers).
Ensure that the packaging will prevent a recontamination of the prod-
uct.
T
t
Water
[°C/°F]
[min]
quality
<25/77
3
D–W
55/131
10
FD–W
>10/50
1
FD–W
90/194
5
FD–W
-
-
-
Chemistry/Note
-
Concentrate, alkaline:
– pH ~ 13
– <5 % anionic surfactant
0.5 % working solution
– pH ~ 11*
-
-
mind. 10 min at max. 120 °C

7.10 Steam sterilization

CAUTION
Damage or destruction of the batteries due to processing!
Do not sterilize the rechargeable battery.
Note
Remove all attached components from the product (tools, accessories)
before sterilization.
Check to ensure that the sterilizing agent will come into contact with
all external and internal surfaces (e.g. by opening any valves and fau-
cets).
Use a validated sterilization method:
– Steam sterilization using fractional vacuum process
– Steam
sterilizer
DIN EN ISO 17665
– Sterilization using fractionated vacuum process at 134 °C/holding
time 5 min
When sterilizing multiple products in one steam sterilizer:
Ensure that the maximum permitted load specified by the manufac-
turer for the steam sterilizer is not exceeded.

7.11 Storage

Store sterile products in germ-proof packaging, protected from dust, in
a dry, dark, temperature-controlled area.
DIN EN 285
and
validated